2025-08-14
Paraffin sectioning in the modern laboratory and the problem of static electricity: a review
Publication
Publication
Journal of Histotechnology , Volume 2025
Build-up of static electricity frequently hampers proper production and handling of serial paraffin sections produced on a microtome, a problem that is exacerbated in modern, fully climatized laboratories. In this context, we review a number of suggested remedies for static electricity, concerning increase of humidity, in one way or the other, and various kinds of anti-static devices. An excellent solution, the shockless one-point ionizing bar that we have implemented in our laboratory, albeit not new, is presented here in detail.
| Additional Metadata | |
|---|---|
| , , , , , , , | |
| doi.org/10.1080/01478885.2025.2544011 | |
| Journal of Histotechnology | |
| Organisation | Staff publications |
|
Paraffin sectioning in the modern laboratory and the problem of static electricity: a review. (2025). Paraffin sectioning in the modern laboratory and the problem of static electricity: a review. Journal of Histotechnology (Vol. 2025). doi:10.1080/01478885.2025.2544011 |
|